1 foot from the unit and again approach the nonadhesive side of the tape with your 
free hand, If the tape has been neutralized, it will not attract. 
 
CAUTION! ELECTRICAL SHOCK HAZARD – Do not insert objects through 
intake or outlet grille. 
   
Do not try to verify operation of the unit by drawing a spark from an ion emitter 
point. The design of the balancing circuit makes the “spark test” inconclusive. 
Sustained grounding of the ion emitters may damage the balancing circuit. 
 
Ion Balance Check 
To test the unit for ion balance, the use of a charged plate monitor such as the 
SIMCO Electrostatic Analyzer H0601(SSD) or cpm374(Warmbier) is recommended. 
Offset voltage should be measured and checked against the Ion Balance in Section 3, 
Specifications. 
Testing value can be changed with difference of temperature, humidity and ambient 
environment. 
Do not try to determine ion balance by holding a static meter in the ionized air 
stream. This will result in s meaningless reading.   
 
 
Ion balance adjustment   
This Blower is designed with function of ion balance adjustment. If the offset 
voltage (ion balance deflection) is too much, you can adjust the ion balance. Use the 
screwdriver to adjust the balance localizers back of the unit respectively. When ion 
balance lean to positive, adjust clockwise. Adjust clockwise to produce negative ion; 
When ion balance leas to negative, adjust anticlockwise. Adjust anticlockwise to 
produce positive ion. 
NOTE: The two fans are adjusted at the same time when adjusting the ion balance. 
 
Calibration 
The BLOWER OVERHEAD’s ion output is inherently balanced by design. If, after 
adjusting the ion balance, the offset voltage is still too much ,please contact our 
Customer Service. 
 

Summary of Contents for OVERHEAD Series

Page 1: ...OVERHEAD is a two Fan unit designed to provide targeted coverage for electrostatic problems MATTERS FOR DAILY INSPECTION Clean high voltage needles Verify grounding connection Inspect equipment for da...

Page 2: ...y The high voltage AC is applied to a circular arrangement of stainless steel ion emitter points which results in an intense alternating electric field at the tip of the emitter points It is this elec...

Page 3: ...ations Unit Part Number OVERHEAD 120VAC OVERHEAD 230VAC Line Voltage 120 VAC 60 Hz 230 VAC 50Hz Current Draw Max 2 Amp 1 Amp Air Volume Output Fan Speed Flow Low 2 35 CFM High 2 70 CFM Air Velocity Fa...

Page 4: ...3 15 300mm Tp2 1 5 tp6 3 5 p10 7 0 tp14 11 0 300mm Tp3 1 5 tp7 3 5 tp11 7 0 tp15 11 0 300mm blower Tp4 5 5 tp8 8 5 tp12 11 0 tp16 15 0 1 Offset voltage and discharge time determined as per ESD Associa...

Page 5: ...The BLOWER OVERHEAD should be placed approximately 1 to 3 feet from the critical work area or objects to be neutralized It should be positioned to cover as much of the area as possible with the ionize...

Page 6: ...he time required to neutralize a static charge on an item in the air stream depends on many factors Two important factors are distance to ionizer and air velocity Air ions constantly neutralize each o...

Page 7: ...is stubborn wash the filter in mild soap and water then rinse Blot the filter dry with paper towels and allow to dry Install filter on air inlet and secure by snapping the filter retainer in place IMP...

Page 8: ...ting value can be changed with difference of temperature humidity and ambient environment Do not try to determine ion balance by holding a static meter in the ionized air stream This will result in s...

Page 9: ...SECTION 8 How to hang...

Reviews: