User manual - RATEMETER STI-94
5. Try to write improper data to register (register 04h):
ADDR
FUNC
REG H,L
DATA H,L
CRC L,H
01
06
00
04
00
10
C9
C7
DATA H, L written value (10h = 16) out of allowable range (0 ÷ 5)
Device response ( with exception code 03h):
ADDR
FUNC
ERR
CRC L,H
01
86
03
09
C1
There is no full implementation of the Modbus Protocol in the device. The
functions presented above are available only.
47