background image

14

voltage must not exceed the Instrument’s rated maximum circuit-to-ground
voltage. (See Table 1-3, item 12.)

16. Do not make electrical measurements where the air may contain explo-

sive concentrations of gas or dust such as in mines, grain elevators, gaso-
line stations or in the presence of charging batteries, until determined to
be safe by qualified personnel. Note that even metallic dust can be ex-
plosive.

17. No general purpose VOM is to be used to make measurements on blast-

ing circuits or blasting caps. Use only designated instruments.

Be extremely careful when working with high voltage circuits. Even though
the Instrument and test leads are well insulated for protection of the operator,
it is not advisable to handle either when power is on in the circuit.

4.2

Measuring DC Voltages

1.

Set the range switch at one of the six voltage range positions marked 0.3
V, 3 V, 12 V, 60 V, 300 V or 600 V. When in doubt about the approximate
voltage present, start with the highest voltage range to protect the Instru-
ment. If the voltage reading is within the limits of a lower range, the switch
then may be set to that range to obtain a more accurate reading.

2.

Be sure the power is off in the circuit to be measured and all capacitors
have been discharged.

3.

Connect the black test lead to the negative side of the circuit being mea-
sured and the red test lead to the positive side of the circuit.

4.

Turn on the power of the circuit and read the voltage on the black scale
marked AC-DC. For the 0.3 V and the 3 V ranges, use the 0-300 figures
and divide the reading by 1000 and 100 respectively. For the 12 V, 60 V
and 300 V ranges read the figures directly. For the 600 V range, use the
0-60 figures and multiply the reading by 10.

5.

Turn the power off, disconnect the test leads and return the range switch
to the OFF position.

4.3

Measuring AC Voltages

NOTE: The rectifier circuit in the 14510-2 responds to the full-wave rectified
average value of an AC waveform. The Instrument is calibrated in terms of the
rms value of a pure sine wave which will be correct for all sinusoidal wave-
forms. If the waveform is non-sinusoidal, the reading could result in a sub-
stantial error. Also, accuracy is lessened at higher input frequencies.

NOTE: Since the VOM will respond to DC voltage when set on any AC volt-
age range, an external blocking capacitor must be employed where mea-
surements of AC superimposed on DC are encountered.

1.

Set the range switch at one of the five voltage range positions marked 3
V, 12 V, 60 V, 300 V or 600 V. When in doubt about the approximate volt-
age in the circuit being measured, start with the highest voltage range as

!

Summary of Contents for 14510-2

Page 1: ...Simpson 14510 2 Volt Ohm Milliammeter OPERATOR S MANUAL...

Page 2: ...rt or Customer Service at 847 697 2260 Internet http www simpsonelectric com Warranty and Returns SIMPSON ELECTRIC COMPANY warrants each instrument and other articles manufactured by it to be free fro...

Page 3: ...nt equipment or general laboratory applications Some high power circuits however are within its measurement range and present an arcing explosion hazard in the event of an unanticipated circuit behavi...

Page 4: ...4 NOTES...

Page 5: ...5 NOTES...

Page 6: ...ts 10 3 CONTROLS JACKS AND INDICATOR 11 3 1 Front Panel 11 4 OPERATION 12 4 1 Safety Precautions 12 4 2 Measuring DC Voltages 14 4 3 Measuring AC Voltages 14 4 4 Measuring Direct Current 15 4 5 Measur...

Page 7: ...n either a vertical or horizontal position The horizontal position is preferred for greater accuracy since the Instrument is calibrated in that posi tion 1 3 Items And Accessories All applicable items...

Page 8: ...atings are 100 V AC or DC or 10 DC amperes These ratings for the test leads must not be exceeded in any application 1 4 Technical Data Table 1 3 lists the Technical Data for the 14510 2 Table 1 3 1 DC...

Page 9: ...sly be applied to the circuits of an instru ment 1 5 Definition Of Accuracy The voltage and current accuracy of the Instrument is commonly expressed as a percent of full scale This should not be confu...

Page 10: ...ligator clips for the test leads are in a polyethylene bag See Section 5 for instructions on how to open the battery compartment and install the bat teries 2 2 Warranty The Simpson Electric Company wa...

Page 11: ...hile others require a multiplying factor 2 Range Switch The range switch has 21 positions It may be turned to any position from either direction There are eleven voltage positions four direct current...

Page 12: ...y be attached to the probe tips Eliminating the need to hand hold test probes to a circuit for extended periods the test clips also reduce hand prox imity to a high voltage circuit while energized Cir...

Page 13: ...certain that someone capable of rendering aid is nearby and aware 8 Do not handle the Instrument its test leads or the circuitry while taking any measurements 9 Be sure to reset the range switch to th...

Page 14: ...ll capacitors have been discharged 3 Connect the black test lead to the negative side of the circuit being mea sured and the red test lead to the positive side of the circuit 4 Turn on the power of th...

Page 15: ...In strument NOTE The voltage drop will not significantly affect most circuits whose cur rent is being measured In some low voltage circuits such as transistor cir cuits it may be necessary to take the...

Page 16: ...rve the Instrument indication It should read 0 on the ohms arc which is at the top of the dial 4 If the pointer does not read 0 rotate the V Adj Zero Adjust Control knob either direction until it does...

Page 17: ...by one of the Authorized Service Centers or by the factory 5 1 Inspection The user is protected from electrical shock by the insulation of the Instrument and the test leads Frequent examination of ea...

Page 18: ...ace the compartment cover If the Instrument still does not operate the internal fuse may be open For access to this fuse the case back must be removed WARNING To avoid electrical shock make certain th...

Page 19: ...ial purpose Replace only with a Littlefuse Type BLS or Bussman type BBS 2 A 600 V 4 Shake the unit gently to make certain that no loose debris is left inside Replace the case back and tighten the four...

Page 20: ...into a warm building Therefore do not attempt high voltage measurements until the Instrument has had time to dry completely 6 SERVICING INSTRUCTIONS An improperly repaired instrument may be dangerous...

Page 21: ...76 R23 Resistor 184 1 1 2 W 1 110271 R24 Resistor 2 02 k 1 1 2 W 1 110270 R25 Resistor 16 8 k 1 1 2 W 1 110292 R26 Resistor 1 21 1 2 Bobbin 10 675499 R27 Resistor 12 22 1 2 Bobbin 10 675201 R28 Resist...

Page 22: ...22 Figure 6 1 Schematic Diagram 14510 2 Volt Ohm Milliammeter 6 2 Schematic Diagram 14510 2Volt Ohm Milliammeter...

Page 23: ...23 NOTES...

Page 24: ...24 SIMPSON ELECTRIC COMPANY 853 Dundee Avenue Elgin IL 60120 3090 847 697 2260 FAX 847 697 2272 Printed in U S A Part No 06 114022 Edition 3 11 02 Visit us on the web at www simpsonelectric com...

Reviews: