The RO membrane housing is located on top of the main system body.
(1)
Remove the locking C-clip from the RO housing end cap QC adaptor.
(2)
Disconnect the PE tube from the QC adaptor.
(3)
Unscrew the housing cap with RO housing wrench.
(4)
Place the new RO membrane into the housing. Two ‘O’-rings first.
(5)
The large black chevron seal should still be visible.
(6)
Push the RO membrane fully into the housing
(7)
Check the ‘O’-ring inside the housing cap is in place and reassemble.
(8)
Tighten the RO housing end cap using the wrench.
(9)
Reconnect the PE tubing and C-clip to the RO housing cap.
Please pay attention to the direction the RO element has to be
installed, or it will not function properly.
•
Pre-Filter Housing Assembly
Now we have the RO membrane element loaded we need to install the
cartridge filters in the three removable housings
Pay special attention to the direction the GAC (UDF)
cartridge is installed, or it may cause a water leak.
19
20
Pre-Filter Housing Bowls
PP Filter
GAC (UDF) Filter
CTO Filter