Press and release , the date and time is set
and the meter turns off.
STEP 8
Note:
Time is displayed in 24 hours format only. If the battery is replaced, it may cause the date and time to
reset.
Resetting the date and time does not change the time stamps of previous test results.
Note:
Unit of Measure is factory preset to mg/dL or mmol/L. It cannot be changed by the user.
15
16
Control Solution Testing
Why Perform a Control Solution Test
•
Ensures that your meter and test strips are working properly.
•
Allows you to practice testing without using blood.
When Should You Perform a Control Solution Test
•
When you begin using a new vial of Test Strips.
•
When the vial cap is opened for a long time.
•
When Test Strips have been exposed to extreme environmental conditions.
•
When you want to check whether meter or test strips are work properly.
•
When you drop or broken the meter.
•
When you want to check whether the test steps are correct.
Important Information
•
Use only Sinocare blood glucose control solution with your Meter.
•
Check the expiration date on the control solution bottle. Do not use if expired.
•
Use the test strips within a period of 6 months from the date that you first open the vial.
•
Record the opened date on the control solution bottle, which is 3 months from when you
first open it. Discard the bottle after 3 months.
•
Discard any bottle that appears to be cracked or leaking.
•
For in vitro diagnostic use only.