39
40
Display
What it means
LED light malfunction
Test result is “HI”
or “LO”, but no red
light on.
Test result is >70mg/dL
(>3.9 mmol/L)and ≤
126 mg/dL(≤ 7.0
mmol/L), but no green
light on
Test result is≤70mg/dL
(≤ 3.9 mmol/L) or >250
mg/dL (>13.9 mmol/L),
but no red light on.
Call Sinocare customer service
at +86-731-89935581/
+86-731-89935582
or contact your local dealers.
What to do
Display
What it means
The test strip package is damaged,
and desiccant tablet becomes soft;
the test strip vial open for too long;
Expired test strip;
Meter displays
incorrect test result
Meter begins to
countdown after
inserting a test strip
without sample
Meter or test strip malfunction
Test strip may be affected with
damp
Test strip inserted incorrectly or
incompletely
Insert the test strips with the arrow
side facing up and in the direction
of the arrows untio it will go no
further.
Use a new test strip to test.
Call Sinocare customer service at
+86-731-89935581/
+86-731-89935582
or contact your local dealers.
Use test strip from a new vial.
Use test strip from a new vial within
USE-BY date.
What to do
If you follow the actions recommended but the problem is not resolved, please contact Customer
Service at +86-731-8986-731-89935582 or contact your local dealers for assistance.