Introduction to the Your Blood Glucose and Uric Acid Monitoring System
Intended Use
The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is designed for the
quantitative measurement of glucose and uric acid in fresh capillary whole blood samples
and in venous whole blood samples. The Safe AQ UG Blood Glucose and Uric Acid Monitor-
ing System is for use outside the body only (in vitro diagnostic use) for self-testing and
professional use as an aid in the management of diabetes and hyperuricemia (HUA).
The Safe AQ UG Blood Glucose and Uric Acid Monitoring System is intended for in vitro
diagnostic use and should not be used for the diagnosis of or screening of diabetes and
hyperuricemia (HUA).
The Safe AQ UG Blood Glucose and Uric Acid Monitoring System includes:
Safe AQ UG blood glucose and uric acid meter, Safe AQ UG blood glucose test strip, Safe AQ
UG uric acid test strip, blood glucose control solution, uric acid control solution
Unit type: Safe AQ UG Blood Glucose and Uric Acid meter displays blood glucose results in
mg/dL or mmol/L and uric acid results in mg/dL or μmol/L.
Test Principle
A blood glucose test or uric acid test is based on measurement of electrical current caused
by the reaction of glucose or uric acid with the reagents (special chemicals) on the
electrode of the strip. The blood or control solution sample is drawn into the tip of the test
strip through capillary action. Glucose or uric acid in the sample reacts with the special
chemicals and generates electrons, which produce electrical current. The meter measures
the electrical current and calculates the glucose or uric acid result. The blood glucose
results are displayed in mg/dL or mmol/L and uric acid results are displayed in mg/dL or μ
mol/L.