16
9. Remove test strip from drop when meter beeps and
begins testing.
10. After testing is finished, result appears in the meter
Display with the Control Symbol.
Compare result to Control Test range printed on test strip vial
label for level of control solution you are using. If result is in
range, system can be used for testing blood. If result does not
fall within range, repeat control test using a new test strip. If
Control Test result is outside range, test again. If result is still
outside range, system should not be used for testing blood.
Contact for assistance (see Booklet cover for contact informa-
tion).
11. After result is shown, remove test strip from meter and
discard. Meter turns off. Recap control solution bottle tightly.
1. Store meter in carrying case to protect from liquids,
dust and dirt. Do not keep meter in an area where it may be
crushed (i.e. back pocket, drawer, bottom of bag, etc.).
2. Clean immediately after getting any blood on the meter
or if meter is dirty or before allowing anyone else to handle it.
Do not use bleach to clean the meter. Make sure meter is off
and a test strip is not inserted. Wipe Meter with a clean, lint
free cloth or swab dampened with 75% alcohol or neutral
cleaner. Let Meter air dry thoroughly before using to test.
6. Preservation and Maintenance