Result Range: 1.1 - 33.3 mmol/L
Sample Size: 0.5 microliter (0.5 μL)
Sample: Fresh capillary whole blood or control solution
Test Time: Results in as fast as 4 seconds
Result Value: Plasma values
Assay Method: Electrochemical
Power Supply: One 3V lithium battery #CR2032 (non-re-
chargeable), Maximum rated current: 10mA
Battery Life: Approximately 1000 tests or 1 year
Manufacture date: Refer to meter label.
Shelf life: 10 years (based on every day perform 7 tests)
Software release version: V01
Memory Size: 500 glucose results
Size: 40mm×36mm×22mm
Operating Range
Relative Humidity: 10%-80% (Non-condensing)
Temperature: 5°C - 40°C
Hematocrit: 20%-70%
Storage condition: 10~93%RH, -20~55℃
Altitude: Up to and including 3108 meters
30
(《China type 2 diabetes prevention guideline》2017)
China type 2 diabetes blood glucose control range
Before meal
4.4~7.0mmol/L
(79 mg/dL~126 mg/dL)
State
Range
Before meal
3.9~6.1 mmol/L
(70 mg/dL~110 mg/dL)
State
Normal range
Gestational diabetes mellitus prevention guideline(2014))
Gestational diabetes mellitus (GDM) glucose control range
Preprandial plasma glucose
(before meal)
≤5.3mmol/L(≤95mg/dL)
State
Range
(National clinical laboratory operating procedures Fourth Edition )
Normal glucose range