background image

Technical Information  

TI0387R9

 

100 HUNTER PLACE, YOUNGSVILLE, NC 27596  USA 

Ph: (919) 554-2244, (800) 356-7311 • Fax: (919) 554-2266, (800) 899-8181  

• Web: www.sirchie.com • Email: sirchieinfo@sirchie.com 

 

Trouble Shooting Guidelines 

Before returning the instrument for repair, carefully review these Trouble Shooting steps to 
determine if the problem can be solved. If the instrument still fails to work correctly, contact 
SIRCHIE at (800) 356-7311 to arrange its return for service. 

 

Symptom 

Possible Causes 

Actions 

Power indicator LED does not 
light; instrument is non-
responsive. 

1) Battery pack not 
charged. 

2) AC adapter connected 
to instrument. 

1) Connect AC adapter to instrument and charge 
battery pack. Refer to Unpacking/Setup for 
instructions. 

2) Remove AC adapter. 

Fault LED lights during charging. 

1) Problem with battery 
pack. 

 

 

 

2) Internal failure. 

1) Disconnect AC adapter from instrument and 
wait one hour for battery pack to cool. Do not 
use instrument or attempt to charge battery 
pack during this time. After one hour, reconnect 
AC adapter to instrument. If red LED (fault) 
illuminates again, replace battery pack. 

2) If battery pack is not the problem, 
discontinue using instrument and return it to 
Sirchie for service. 

Instrument does not seem 
operable; does not alarm in the 
presence of leaks. 

1) Sensor failure. 

2) Instrument failure. 

 

 

Attempt to verify unit operation on known 
combustible gas leak source. If no response, 
return the instrument to Sirchie for service. 

 

Battery life is less than3 hours of 
operation. 
 

1) Battery pack needs to 
be conditioned. 
 

1) To condition battery pack: charge it fully, then 

operate instrument as normal until the low 
battery LED illuminates. Repeat this procedure a 
minimum of 3 to 5 times. 

Instrument does not operate 
and battery pack charge cycle 
will not complete (24+ hour 
charge applied and green light 
does not illuminate). 

1) Battery pack not 
connected. 
 
2) Bad battery pack. 

1) Follow instructions in the Unpacking

/

Setup 

section to check that the battery pack 
connection is secure. 

2) Replace battery pack. 

CAUTION: This instrument should be serviced only by SIRCHIE. Failure to do so may impair the 
intrinsic safety of the device. Do not open the case of the instrument. There are no user-
serviceable parts inside.  

Only the battery compartment section of the case may be opened to insert or remove the 
battery pack. 

Reviews: