8 Preparing the Chair for Installation
Sirona Dental Systems GmbH
59 58 470 D 3370
12
D 3370.031.01.10.02
1.
Remove the front faceplate of the base.
It is fitted on the lateral faceplates (if necessary, careful-
ly prize the faceplates apart with a screwdriver).
2.
Unscrew the two screws (with washers) at the front of
the cover. Take off the cover.
3.
Remove the transport safety device.
4.
Pull off the small bellows from the foot control.
Remove the rear faceplate of the base. It is fitted on the
lateral faceplates (if necessary, carefully prize the face-
plates apart with a screwdriver).
5.
Unscrew the top faceplate (two screws with washers).
6.
Loosen the bellows in the lower region.
7.
Separate the chair from the pallet:
Unscrew four fastening screws (with washers).
8.
Remove lateral cover on each side.
NOTE
i
If the C2+ is being assembled as a replacement for an M1
that was previously located at this position, an adapter
board can be ordered (Order No. 58 70 493) with which the
existing mounting holes can still be used. For installation in-
formation regarding this point refer to “Fastening the Chair
to the Floor with the M1 Adapter Plate” on page 42
NOTE
i
A chair plate can be ordered for particularly uneven floors
(Order No. 58 71 913) which compensates the unevenness
of the floor. For installation information regarding this point
refer to “Fastening the Chair Base to the Floor” on page 13
8 Preparing the Chair for Installation
1.
2.
4.
5.
6.
7.
7.
8.
3.