Sirona Dental Systems GmbH
17 Water Unit: Making Connections
59 58 470 D 3370
D 3370.031.01.10.02
23
båÖäáëÜ
NOTE
i
If the drain pipe is in an unfavorable position (such as termi-
nation panel M1), the drain hose can be moved as illustrat-
ed.
1.
Connect both connector pieces and supply hoses to-
gether and secure them with a clamp.
NOTE
i
The fronts of the connector pieces must be firmly enclosed
by the clip and the clip must engage.
2.
Connector
X33
:
Insert cable sleeve of electrical lead into correspond-
ingly marked recess of terminal box. Plug in connector.
3.
Connector
X7
and cable
X1
:
Plug connector
X7
into fuse board
CF
.
Connect gray (GY) cable onto
X1.1
(board CF) and pink
(PK) cable onto
X1.2
.
ATTENTION
Make sure you connect the cables the right way around.
4.
Insert cable through recess in fuse box and fasten with
washer
A
.
5.
Remove connector housing
X8
(jumper with marking
tag) from board
CJ
. Plug in connector housing
X8
of
cable
L32
in its place.
6.
Plug connector
X99
of cable L43 to board CJ.
17 Water Unit: Making Connections
A
X7
CF
X1
PK
GY
X2
X8
X1
X7
X3
1
2
Grease SYNTHESO PRO AA2
1.
2. / 6.
3.
4.
X8
X99
X33
i
5.
1
70
mm±1
0
100mm
20 41 648
10
+
_
!
54 44 018
29 40 120