Sirona Dental Systems GmbH
36 Reversing the Direction of the Travel Track
59 58 470 D 3370
D 3370.031.01.10.02
49
båÖäáëÜ
ATTENTION
Unplug the unit!
1.
Pull off and swap the positions of the connectors for the
white and brown leads at the travel track drive (white in
place of brown and brown in place of white).
2.
Pull off socket casing
R6
from the reversing block.
3.
Press lightly on spring clamp
A
on
R6.1
and
R6.2
and
detach the cores.
4.
Press the spring clamps up slightly and attach the
cores again in reverse (white to
R6. 2
and green to
R6.1
).
5.
Attach socket housing
R6
again.
36 Reversing the Direction of the Travel Track
1
2
3
A
R6.2
R6.1
R6
1.
2.
3.
4.
5.
br