Sirona Dental Systems GmbH
19 Assistant Element: Making Hose Connections
59 58 470 D 3370
D 3370.031.01.10.02
25
båÖäáëÜ
Version with and without disinfection system
1.
Pull the nipple holder off from the four-way manifold and
plug in the water hose
W38
from the assistant element.
Refit the nipple holder.
2.
Pull the nipple holder off from the three-way manifold
and plug in the air hose
A39
from the assistant element.
Refit the nipple holder.
ATTENTION
Make sure that the nipple holder is firmly seated in the guide
grooves of the hose nipples. Check by pulling on the nip-
ples.
Stow the manifolds in the rear bottom area
.
3.
Attach the nipple of the hose
W43
to its matching coun-
terpart.
Snap on the safety clip.
4.
Attach the nipple of the hose
A34
to its matching coun-
terpart.
Snap on the safety clip.
NOTE
i
Step
4.
also applies to
MV33
in combination with a wet suc-
tion system.
5.
Arrange the hose connections below the fuse box as il-
lustrated.
19 Assistant Element: Making Hose Connections
1.
2.
4.
A41
A41
W32
W32
5.
3.
A39
W38
W43
W43
A34
A34