31 Fastening the Chair to the Floor with the M1 Adapter Plate
Sirona Dental Systems GmbH
59 58 470 D 3370
42
D 3370.031.01.10.02
1.
Align the adapter plate with the M1 connection panel.
2.
Use the existing drill holes (3) to secure the adapter
plate.
– With a
concrete floor
:
Use 10 x 160 wood screws and washers.
– With a
wooden floor
:
Use 10 x 80 wood screws and washers.
3.
Grasp the chair in the area near
M
, lift it from the pallet
and position it on the adapter plate.
ATTENTION
Do not lift the chair by its backrest or by the bracket for the
upholstery.
4.
Fasten the chair to the adapter plate at points
A
,
B
,
D
,
E
and
K
with the M10 x 18 screws.
NOTE
i
The chair must be fastened at positions
C
and
H
, after the
electrical connection has been made. The chair can then be
moved upward with the foot control (refer to page 18 ).
5.
Screw protective ground conductor
S
to the M1 adapter
plate and the bottom of the chair (do not forget contact
plate
KS
).
•
Continue with installing the chair at the section entitled
“Connecting Air and Water” on page 15