Sirona Dental Systems GmbH
1 Safety
59 58 470 D 3370
D 3370.031.01.10.02
5
båÖäáëÜ
ATTENTION
It is essential that you comply with the warning and safety
information contained in the Installationsanleitung.
All such information is highlighted by the captions NOTE,
ATTENTION, and CAUTION.
ATTENTION
The installation must be carried out in accordance with the
requirements stated in our Installationsanleitung.
ATTENTION
Installation may be carried out only by personnel expressly
authorized by Sirona.
ATTENTION
It is essential that you switch off the treatment center PRIOR
to starting the installation.
ATTENTION
For reasons of product safety, only original Sirona accesso-
ries approved for this product, or accessories from third par-
ties approved by Sirona, may be used. The user is respon-
sible for dangers resulting from the use of non-approved
accessories.
If any devices not approved by Sirona are connected, they
must comply with the applicable standards, e.g.:
• IEC 60950 for information technology equipment and
• IEC IEC 60601-1 for medical electrical equipment
In case of doubt, contact the manufacturer of the system
components.
ATTENTION
Any person who assembles or modifies a medical electrical
system complying with the standard IEC 60 601-1-1 (safety
requirements for medical electrical equipment) by combin-
ing it with other equipment (e.g. when connecting a PC) is
responsible for ensuring that the requirements of this regu-
lation are met to their full extent for the safety of the patients,
the operators and the environment.
ATTENTION
To ensure the operational readiness of medical electrical
equipment, the use of mobile radio telephones in the prac-
tice or hospital area is prohibited.
ATTENTION
Use an ESD wrist band.
Connect it to the
protective ground wire.
ATTENTION
When opening the unit:
Please observe the safety measures for han-
dling PC boards (ESD).
Touch a ground point to discharge static elec-
tricity before handling any components.
1 Safety