2 Dimensions, technical data
Sirona Dental Systems GmbH
Installation Requirements C5
+
, C5
+
Turn
60 27 978 D3415
28
D3415.021.03.08.02
2.3 Technical data
Technical data
C5
+
,
C5
+
Turn
Power supply line
100 V / 115 V / 230 VAC
50 / 60 Hz
Nominal current
11.5 A
at 100 V AC
9,5 A
at 115 V AC
4,5 A
at 230 V AC
In addition, max. 6 A for external devices
Main unit fuse
at 100/115 V AC
T 10 A H, 250 V AC , REF 10 77 460
at 230 V AC
T 6,3 A H, 250 V AC , REF 10 77 452
Transport and storage conditions
Temperature:
-40°C – +70°C (-40°F – +158°F)
Relative humidity:
10% – 95%
Air pressure:
500hPa – 1060hPa
Operating conditions
Ambient temperature:
10°C – 40°C (50°F – 104°F)
Relative humidity:
30% – 85% (no condensation)
Air pressure:
700hPa – 1060hPa
Maximum load capacity of the patient chair
135kg, including accessories
Installation site
≤
3000 m above sea level
Pollution degree
2 acc. to IEC 60664-1
Protection class
Class I device
Device class in accordance with Directive
93/42/EEC
Class IIa equipment
Degree of protection against electric shock
Type B applied parts,
except SIROCAM 3/SIROCAM C:
Type BF applied parts
Degree of protection against ingress of water
Ordinary equipment (not protected)
The foot switch has an IPX1 degree of protection against liquids
(drip-proof).
Operating mode
Continuous operation with intermittent loading corresponding to the
dental mode of working.
Permanently connected unit.
Year of manufacture
(on rating plate of chair)
On-site pressure readings (1 bar = 100 kPa = 14.5 psi)
Air min./max.
Water min./max.
550 kPa / 750 kPa (5.5 / 7.5bar)
250 kPa / 600 kPa (2.5 / 6bar)
Foot control wireless interface (not possible for C5
+
Turn)
Model designation:
Frequency:
Transmitting power:
Modulation type:
Range:
nanoLOC AVR
2,4 GHz – 2,4835 GHz (ISM band)
< 2 mW (short range device)
MDMA
approx. 10 m