3 Electromagnetic compatibility 

Sirona Dental Systems GmbH

Installation Requirements C5

+

, C5

+

Turn

60 27 978 D3415

36

D3415.021.03.08.02

båÖäáëÜ

The 

UNIT

 is intended for operation in the electromagnetic 

environment specified below. 

The customer or user of the 

UNIT

 should make sure that it 

is used in such an environment.

3.3 Immunity to interference

Immunity interfer-
ence tests

IEC 60601-1-2 test level

Conformance level

Electromagnetic environment 
guidelines

Electrostatic dis-
charge (ESD) accord-
ing to 

IEC 61000-4-2

± 6kV contact discharge

± 8 kV air discharge

± 6kV contact discharge

± 8kV air discharge

Floors should be made of wood or 
concrete or covered with ceramic til-
ing. If the floor surface consists of 
synethetic material, the relative 
humidity must be at least 30%.

Electrical fast tran-
sient/burst 
according to 

IEC 

61000-4-4

± 1kV for input and output 
lines

± 2kV power cables

± 1kV for input and output 
lines

± 2kV power cables

The quality of the supply voltage 
should conform to the typical busi-
ness or hospital environment.

Surge voltages 
according to 

IEC 

61000-4-5

± 1kV push-pull voltage

± 2kV push-pull voltage

± 1kV push-pull voltage

± 2kV push-pull voltage

The quality of the supply voltage 
should conform to the typical busi-
ness or hospital environment.

Voltage dips, short 
interruptions and vari-
ations of the power 
supply 
according to

 IEC 

61000-4-11

<5% U

T

 for ½ period (>95% 

dip of U

T

)

40% U

T

 for 5 periods (60% 

dip of U

T

)

70% U

T

 for 25 periods (30% 

dip of U

T

)

<5% U

T

 for 5sec. (>95% dip 

of U

T

)

<5% U

T

 for ½ period 

(>95% dip of U

T

)

40% U

T

 for 5 periods (60% 

dip of U

T

)

70% U

T

 for 25 periods 

(30% dip of U

T

)

<5% U

T

 for 5sec. 

(>95% dip of U

T

)

The quality of the supply voltage 
should correspond to the typical busi-
ness or hospital environment.

If the user of the 

UNIT

 requires it to 

continue functioning following inter-
ruptions of the power supply, it is rec-
ommended to have the 

UNIT

 

powered by an uninterruptible power 
supply or a battery.

Magnetic field of 
power frequencies 
(50/60 Hz) 
according to 

IEC 

61000-4-8

3 A/m

3 A/m

The power frequency magnetic fields 
should correspond to the typical val-
ues found in the relevant business 
and hospital environment.

Remarks: U

T

 is the AC supply voltage prior to application of the test level.

Summary of Contents for C5+ Turn

Page 1: ...kÉï ë çÑW NMKOMNQ båÖäáëÜ RH I RH qìêå fåëí ää íáçå oÉèìáêÉãÉåíë ...

Page 2: ...1 3 Media quality 8 2 4 Standards Approvals 30 About this document This document describes the installation requirements for the C5 C5 Turn dental treatment centers The following sets of Installation Instructions describe the subsequent installation procedure REF 59 57 472 C5 REF 59 74 675 C5 Turn You will also need the following drilling template in order to fasten the treatment center securely t...

Page 3: ...cuspidor not for C5 Turn 13 1 7 Mounting plates for equipment with swiveling cuspidor not for C5 Turn 14 1 8 M1 adapter plate for equipment with swiveling cuspidor not for C5 Turn 15 1 9 Demonstration chair plate C5 Turn 16 2 Dimensions technical data 17 2 1 Dimensions of the C5 1 20 18 2 2 Dimensions of the C5 Turn 1 20 21 2 3 Technical data 28 2 4 Standards Approvals 30 3 Electromagnetic compati...

Page 4: ...List of Contents Sirona Dental Systems GmbH Installation Requirements C5 C5 Turn 60 27 978 D3415 4 D3415 021 03 08 02 ...

Page 5: ...ly lines in the termination panel 9 1 5 Underfloor installation of SIVISION connections 12 1 6 Mounting plates for equipment with fixed cuspidor not for C5 Turn 13 1 7 Mounting plates for equipment with swiveling cuspidor not for C5 Turn 14 1 8 M1 adapter plate for equipment with swiveling cuspidor not for C5 Turn 15 1 9 Demonstration chair plate C5 Turn 16 ...

Page 6: ...he user is responsible for any damage re sulting from the use of non approved accessories If any equipment not approved by Sirona is con nected it must comply with the applicable stan dards e g IEC 60950 1 for information technology equip ment and IEC 60601 1 for medical electrical equipment In case of doubt contact the manufacturer of the system components WARNING Any person who assembles or modi...

Page 7: ... of drinking water It can be connected directly to the public drinking water supply The treatment center therefore also bears the symbol for the German Gas and Water Association DVGW along with the rating plate Treatment center without isolation from public drinking water supply If compliance with EN 1717 is stipulated nationally appro priate equipment must be installed beyond the treatment center...

Page 8: ...nnection must be made to cold water When laying the water pipe to the treatment center comply with the following instructions to reduce the quantity of micro organisms in the feed pipe Avoid long stub lines to the treatment center Carry out the installation so that where possible other main consumers such as the sink are fed from the same line downstream of the treatment center connection Avoid la...

Page 9: ...8 D3415 D3415 021 03 08 02 9 1 4 Supply lines in the termination panel 5 3 6 6 6 7 Center base plate opening to floor 1 2 4 9 1 3 8 min 500mm 20 7 6 max 60mm 2 3 8 3 4 5 max 5 mm 3 16 max 5 mm 3 16 3 8 1 2 10 x 1mm 10 x 1mm 3 8 x 1 32 3 8 x 1 32 1 3 8 Foot end 3 6 5 m m 5 0 m m 3 6 5 m m ...

Page 10: ...15 830 If necessary you can also prepare the template your self based on the above sketch not true to scale Supply lines in the termination panel Table 1 Supply lines Item Description 1 Water inlet pipe 10x1 mm corner valve outlet 3 8 2 Compressed air inlet pipe 10x1mm corner valve outlet 3 8 3 Suction pipe DN40 HT PP DIN 19560 polypro pylene inner diameter 36 5mm 4 Water drain DN40 HT PP DIN 1956...

Page 11: ...hly metal chips Supply through the floor underfloor installation 1 The top edges of the corner valves for air and water must not protrude more than 60mm above the upper surface of the finished floor 2 The suction and drain pipes must be flush with the up per surface of the floor a deviation of 5mm is permis sible Internal diameter for both pipes 36 5mm 3 The electrical cables must protrude at leas...

Page 12: ...L38 are not yet cut to length and terminated on the PC side It would be impossi ble to pull the cables through when installing them under floor level if a sub D connector were already connected These cables should always be pulled Free length A of cables at the treatment center end Length A 600mm If S video cable L15 is equipped with both a female and a male connector make sure that the female con...

Page 13: ... Turn Mounting plate Demonstration chair plate REF 58 26 511 REF 47 08 074 For especially uneven floor conditions a steel mounting plate is available to compensate for the irregularities of the floor REF 58 26 511 For floors which do not permit permanent connection of the unit e g demo operation at a fair installation on a steel demonstration chair plate is possible REF 47 08 074 ...

Page 14: ...C5 Turn Mounting plate Demonstration chair plate REF 58 71 913 REF 58 66 301 For especially uneven floor conditions a steel mounting plate is available to com pensate for the irregularities of the floor REF 58 71 913 For floors which do not permit permanent connection of the unit e g demo operation at a fair installation on a steel demonstration chair plate is possible REF 58 66 301 ...

Page 15: ...1 03 08 02 15 1 8 M1 adapter plate for equipment with swiveling cuspidor not for C5 Turn Adapter plate If the C5 is installed as a replacement for an M1 at this position then an adapter plate with the aid of which the existing mounting holes can be used is available under the REF 58 70 493 REF 58 70 493 ...

Page 16: ...15 16 D3415 021 03 08 02 1 9 Demonstration chair plate C5 Turn Demonstration chair plate 750 1266 0 5 For floors which do not permit per manent connection of the unit e g demo operation at a fair installa tion on a steel demonstration chair plate is possible REF 59 46 269 REF 59 46 269 C5 Turn ...

Page 17: ...5 C5 Turn 60 27 978 D3415 D3415 021 03 08 02 17 båÖäáëÜ 2 Dimensions technical data 2 1 Dimensions of the C5 1 20 18 2 2 Dimensions of the C5 Turn 1 20 21 2 3 Technical data 28 2 4 Standards Approvals 30 ...

Page 18: ...e C5 1 20 Recommended distances from cabinet or wall Hazard warning The here mounted lamp the dentist element with tray option and the assistent element with long support arm have a wider swivelling range than the distances indicated Center floor opening 250 9 7 8 1100 43 5 16 1100 43 5 16 2500 98 7 16 350 13 3 4 550 21 5 8 with Swivellable cuspidor ...

Page 19: ...D3415 D3415 021 03 08 02 19 båÖäáëÜ Standard upholstery max backrest and chair in highest position min backrest and chair in lowest position 1212 47 3 4 max 1340 52 3 4 min 810 31 7 8 max 760 30 min 400 15 3 4 45 1 3 4 1870 73 3 4 140 5 1 2 300 11 3 4 with Swivellable cuspidor SIROLUX FANTASTIC LEDview ...

Page 20: ... 021 03 08 02 180 7 250 9 7 8 750 29 1 2 770 30 3 8 45 1 3 4 max 1500 59 max 1515 59 5 8 180 315 max 1200 47 1 4 Option Tray ca 830 32 5 8 2070 50 81 1 2 2 max 1200 47 1 4 440 17 3 8 max 380 15 15 5 8 430 15 17 5 8 Option Support arm long Option Support arm short Option Swivellable cuspidor according to chair height ...

Page 21: ...3415 021 03 08 02 21 båÖäáëÜ 2 2 Dimensions of the C5 Turn 1 20 250 9 7 8 1600 63 1600 63 2500 98 7 16 350 13 3 4 Recommended distances from cabinet or wall Hazard warning The here mounted lamp the dentist element with tray option have a wider swivelling range than the distances indicated Center floor opening ...

Page 22: ... 978 D3415 22 D3415 021 03 08 02 250 9 7 8 1050 41 3 8 1050 41 3 8 2700 106 1 4 350 13 3 4 1600 63 Minimum clearances for conversion of treatment center from right handed to left handed version an vice versa 2700x2650mm 1084 42 3 4 1600 63 Cabinet right or left max 900mm height Center floor opening ...

Page 23: ...0 27 978 D3415 D3415 021 03 08 02 23 båÖäáëÜ 1050 41 3 8 1300 51 1 8 250 9 7 8 3650 143 3 4 350 13 3 4 2034 80 1250 49 1 4 Minimum clearances for conversion of treatment center from right handed to left handed version an vice versa 3650x2350mm Storage space max 900mm height Center floor opening ...

Page 24: ...ation Requirements C5 C5 Turn 60 27 978 D3415 24 D3415 021 03 08 02 250 9 7 8 1050 41 3 8 1050 41 3 8 2700 106 1 4 350 13 3 4 1084 42 3 4 1600 63 Equal working conditions for right handers and left handers Position for right handers 1600 63 17 Center floor opening ...

Page 25: ...n Requirements C5 C5 Turn 60 27 978 D3415 D3415 021 03 08 02 25 båÖäáëÜ 250 9 7 8 1050 41 3 8 1050 41 3 8 2700 106 1 4 350 13 3 4 1600 63 Equal working conditions for right handers and left handers Position for left handers 1600 63 1084 42 3 4 17 Center floor opening ...

Page 26: ...5 C5 Turn 60 27 978 D3415 26 D3415 021 03 08 02 SIROLUX FANTASTIC LEDview Standard upholstery max backrest and chair in highest position min backrest and chair in lowest position 1266 49 7 8 max 1340 52 3 4 min 810 31 7 8 max 775 30 1 2 min 415 16 3 8 45 1 3 4 1870 73 3 4 300 11 3 4 ...

Page 27: ...ments C5 C5 Turn 60 27 978 D3415 D3415 021 03 08 02 27 båÖäáëÜ 180 7 250 9 7 8 750 29 1 2 770 30 3 8 max 1500 59 180 315 max 1200 47 1 4 Option Tray ca 830 32 5 8 210 45 1 3 4 max 380 15 15 5 8 2070 50 81 1 2 2 800 50 31 1 2 2 440 17 3 8 430 15 17 5 8 according to chair height ...

Page 28: ...Installation site 3000 m above sea level Pollution degree 2 acc to IEC 60664 1 Protection class Class I device Device class in accordance with Directive 93 42 EEC Class IIa equipment Degree of protection against electric shock Type B applied parts except SIROCAM 3 SIROCAM C Type BF applied parts Degree of protection against ingress of water Ordinary equipment not protected The foot switch has an I...

Page 29: ... 28cm Weight C5 1kg 2 2lbs incl without packaging Dentist element assistant element water unit Chair Upholstery SIROLUX LEDview 37kg 13kg 44kg 16 5kg 126kg 96kg 13kg 10kg 14kg 10kg Dimensions of the packaging C5 Turn Dentist element assistant element water unit Chair Upholstery SIROLUX LEDview 120cm x 63cm x 176cm 157cm x 65cm x 83cm 120cm x 52cm x 40cm 10 5cm x 56cm x 28cm Weight C5 Turn 1kg 2 2l...

Page 30: ...adian Standard Association CSA in accordance with CAN CSA C22 2 No 60601 1 The amalgam separator achieves a separation efficiency of 95 The unit thus fulfills the requirements of ISO 11143 Separating procedure type 1 Centrifugal system The amalgam separator is approved by the German Institute for Struc tural Engineering DIBT and by the French International Organization for Standardization AFNOR Wh...

Page 31: ...E directive 1999 5 EC Standards EN 60950 1 EN 301489 1 EN 301489 17 EN 300328 The modules meet the requirements of the Federal Communications Commission Part 15 of the FCC Rules FCC ID SIFNANOLOCAVR0108 Industrie Canada The modules meet the requirements of Industry Canada RSS2210 IC 7654A nanoLOCAVR The current approvals for the foot control are listed on the rating label on the underside of the p...

Page 32: ...2 Dimensions technical data Sirona Dental Systems GmbH Installation Requirements C5 C5 Turn 60 27 978 D3415 32 D3415 021 03 08 02 ...

Page 33: ... interference 36 3 4 Working clearances 38 3 5 Foot control wireless interface 39 IMPORTANT i The C5 C5 Turn fulfills all requirements for electromagnetic compatibility EMC compliant with IEC 60601 1 2 The C5 C5 Turn is referred to as UNIT in the following Observance of the following information is necessary to ensure safe operation regarding EMC aspects ...

Page 34: ... devices If this proves to be unavoidable the UNIT should be monitored to check and make sure that it is used properly The EMC measurements were performed with the following PC 3 1 Accessories Making the PC connection Designation of interface cables to the PC Supplier XGA cable 10m Sirona S video cable 10m Sirona RS232 cable 10m Sirona 2ndprotectivegroundwire 2 5mm2 10m Sirona PC as peripheral dev...

Page 35: ...ic environment guidelines HF emission according to CISPR 11 Group 1 The UNIT uses HF energy only for its internal function The HF emission is therefore very low and it is improbable that nearby electronic devices might be disturbed HF emission according to CISPR 11 Class B The UNIT is intended for use in all facilities including residen tial areas and in any facilities connected directly to a publ...

Page 36: ... ness or hospital environment Surge voltages according to IEC 61000 4 5 1kV push pull voltage 2kV push pull voltage 1kV push pull voltage 2kV push pull voltage The quality of the supply voltage should conform to the typical busi ness or hospital environment Voltage dips short interruptions and vari ations of the power supply according to IEC 61000 4 11 5 UT for period 95 dip of UT 40 UT for 5 peri...

Page 37: ... in the vicinity of equipment bearing the following graphic symbol a The higher frequency range applies at 80MHz and 800MHz b The field strength of stationary transmitters such as the base stations of radio telephones and land mobile services amateur radio stations as well as AM and FM radio and television broadcasting stations cannot be accurately predeter mined An investigation of the location i...

Page 38: ...e as specified above For transmitters whose maximum nominal output is not specified in the above table the recommended working clearance d in meters m can be determined using the equation in the corresponding column where P is the max imum nominal output of the transmitter in watts W speci fied by the transmitter manufacturer Annotation 1 The higher frequency range applies at 80 MHz and 800 MHz An...

Page 39: ...he treatment center These modules transmit the foot control signals CAUTION Interference with the wireless transmission This wireless transmission may cause interference with or be disturbed by other radio services Wireless module in the wireless foot control and in the treatment center Fußschalter Funkschnittstelle 3 5 Foot control wireless interface Model designation nanoLOC AVR Frequency 2 4 GH...

Page 40: ...póëíÉãë dãÄe OMMNJOMNQ péê ÅÜÉW ÉåÖäáëÅÜ mêáåíÉÇ áå dÉêã åó aPQNRKMONKMPKMUKMO NMKOMNQ ûKJkêKW NNV MTQ fãéêáã Éå ääÉã ÖåÉ páêçå aÉåí ä póëíÉãë dãÄe áå íÜÉ rp W c Äêáâëíê É PN SQSOR _ÉåëÜÉáã dÉêã åó ïïïKëáêçå KÅçã páêçå aÉåí ä póëíÉãë ii QUPR páêçå aêáîÉI pìáíÉ NMM Ü êäçííÉI k OUOTP rp lêÇÉê kç SM OT VTU aPQNR ...

Reviews: