Sirona Dental Systems GmbH
1 Preparations
Installation Requirements C8
+
60 28 000 D3417
D3417.021.04.09.02
11
båÖäáëÜ
Important information for the installer
Depending on the prevailing local conditions, the existing
cable set can be installed in the cable duct of an underfloor
installation by an installer prior to the installation of the den-
tal treatment center. In this case, please observe the follow-
ing:
Proceed with
extreme care
when running the cables. Par-
ticularly cables
L15
and
L38
are very sensitive, and must
never be kinked or twisted. The cables must
not overlap or
cross one another
.
RS232 (
L37
) and the XGA cable (
L38
) are not yet cut to
length and terminated on the PC side. It would be impossi-
ble to pull the cables through when installing them under
floor level if a sub D connector were already connected.
These cables should always be pulled.
Free length
A
of cables at the treatment center end:
Length
A
= 600mm
If S video cable
L15
is equipped with both a female and a
male connector, make sure that the female connector
(socket) points to the connection box of the treatment
center.
1.
Bend the wire at the front end of cables
L37
and
L38
to
form a hook.
2.
Pull cables
L15, L37
and
L38
as well as the
audio
ca-
ble and protective ground wire from the treatment cent-
er through the cable duct to the location of the SIVI-
SION PC.
Save the accessory parts for final installation!
1.4 Underfloor installation of SIVISION connections
2.
1.
A
L15, L37, L38, Audio
protective ground wire
SIVISION
Summary of Contents for c8+ turn
Page 1: ...kÉï ë çÑW MVKOMMV båÖäáëÜ UH fåëí ää íáçå oÉèìáêÉãÉåíë ...
Page 4: ......
Page 48: ......
Page 55: ......