3 Electromagnetic compatibility
Installation Requirements C8
+
60 28 000 D3417
D3417.021.04.09.02
51
båÖäáëÜ
The
UNIT
is intended for operation in the electromagnetic
environment specified below.
The customer or user of the
UNIT
should make sure that it
is used in such an environment.
3.2 Electromagnetic emission
Emission measurement
Conformity
Electromagnetic environment guidelines
HF emission according to
CISPR 11
Group 1
The
UNIT
uses HF energy only for its internal function. The
HF emission is therefore very low, and it is improbable that
nearby electronic devices might be disturbed.
HF emission according to
CISPR 11
Class B
The
UNIT
is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
dential purposes.
Harmonics
according to
IEC 61000-3-2
Class A
Voltage fluctuations / Flicker according
to
IEC 61000-3-3
compliant