Code no.
Description and User Instruction
L
62: The right temperature for building up pressure
for sterilization has not been reached within
7.5 min.
Increase the “Ambient sensitivity” level settings
→
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the
water level is too low
→
If the autoclave is placed in a cold spot remove it to surroundings at room temperature
→
Press C to reset
and then restart the cycle
User manual, Sec.4
64: The safety system has been activated
Press C
→
Restart the autoclave.
User manual, Sec.4
67: The time of sterilization has been less than 3 min.
The instruments should be sterilized again
User manual, Sec.4
70: The pressure level in the chamber does not decrease
fast enough after pressure test
Empty the chamber for water
→
Clean or change the filter
→
Check that the drain tube is not blocked or broken
→
Empty the drain tank
if full
User manual, Sec.4
Video 70
71: The autoclave blocks due to use of bad water
Empty the water tank if necessary
→
Pour in fresh demineralized or distilled water (below 3.0µS/cm)
User manual, Sec.4
Video 71
72: The closing system was blocked as the lid was on its
way up
Remove possible blocking of the lid
User manual, Sec.4
73: The sterilization temperature has not been
reached fast enough during the back-flush
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the water level is too low
→
Press C to reset and then
restart the cycle
User manual, Sec.4
74: The right temperature for building up pressure
for sterilization has not been reached fast
enough
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the water level is too low
→
Increase the “Ambient
sensitivity” level settings
→
Press C to reset and then restart the cycle
User manual, Sec.4
75: The right sterilization temperature has not
been reached fast enough
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the water level is too low
→
Increase the
“Ambient sensitivity” level settings
→
Press C to reset and then restart the cycle
User manual, Sec.4
81: Low level in either the water tank or the oil
bottle
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank or change the oil bottle
User manual, Sec.4
Video 81
84: The sterilization temperature has not been
maintained during sterilization
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if needed
User manual, Sec.4
Video 84
86: The safety ring has been activated
Press C to reset the autoclave and then restart the machine
→
Make sure that the handle is correctly placed in the holder
User manual, Sec.4
Video 86
88: The steam pressure is too low during
sterilization
Pour demineralized or distilled water (below 3.0µS/cm) into the water tank in case of low water level
→
Press C to reset the
autoclave and then restart the cycle
User manual, Sec.4
89: The steam pressure is too high during
sterilization
Check that the drain tube is not blocked, broken or bent
→
Empty the drain tank if full
→
Press C to reset the autoclave and
then restart the cycle
User manual, Sec.4
90: The pressure during pressure test was too low
Check that your compressor is turned on
→
Check the gasket on the lid
→
Remove potential dirt
→
Make sure that the lid is
placed correctly in the lid holder
Check if the O-ring for the filter house is mounted correctly
→
Change the O-ring if damaged
→
Mount the lid for the filter
house correctly
→
Replace the gasket on the lid if it is broken
User manual, Sec.4
Video 90
91: The pressure fell during the pressure test
Check the gasket on the lid
→
Remove potential dirt
→
Make sure that the lid is placed correctly in the lid holder
Check if the O-ring for the filter house is mounted correctly
→
Change the O-ring if damaged
→
Make sure that the lid is
placed correctly in the holder
→
Mount the lid for filter house correctly
→
Replace the gasket on the lid if it is broken
User manual, Sec.4
Video 91
92: The power supply to the autoclave has been
cut out, or the C button has been pressed
during the cycle
Press C to reset the machine and then restart the cycle
User manual, Sec.4
Video 92
94: Oil consumption during standby
Disconnect power from the autoclave and wait 10 seconds
→
Reconnect the autoclave to the power and then restart
User manual, Sec.4
95: This error can only occur if the autoclave is
connected to an automatic water supply. The
autoclave has “asked” for water more than 7
times during standby
Disconnect power from the autoclave and wait 10 sec.
→
Reconnect the autoclave to the power and then restart
→
Check
that water is being supplied to the autoclave
→
Check if the autoclave has a leakage.
User manual, Sec.4
97: No water during external wash
Press C to reset the machine
→
Verify that the chamber is drained. If not, empty the chamber (see section 3.5 Drain Chamber
in the User manual)
→
Restart the cycle
User manual, Sec.4
98: Lid did not close
Press C to reset and wait until the lid is in top position
→
Make sure that the lid is placed correctly in the lid holder (look for
things that might obstruct complete closure and remove these)
→
Start the machine again
User manual, Sec.4
If the above does not solve the problem or for other possible
displayed messages, please refer to the user manual.
Alternatively, please contact your authorized supplier.