Code no. 

Description and User Instruction 

62: The right temperature for building up pressure 

for sterilization has not been reached within 

7.5 min.

 

Increase the “Ambient sensitivity” level settings 

→ 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the 

water level is too low 

→ 

If the autoclave is placed in a cold spot remove it to surroundings at room temperature 

→ 

Press C to reset 

and then restart the cycle 

User manual, Sec.4 

64: The safety system has been activated

 

Press C 

 Restart the autoclave.

 

User manual, Sec.4 

67:  The time of sterilization has been less than 3 min.

 

The instruments should be sterilized again

 

User manual, Sec.4 

70: The pressure level in the chamber does not decrease 

fast enough after pressure test

  

Empty the chamber for water 

→ 

Clean or change the filter 

→ 

Check that the drain tube is not blocked or broken 

→ 

Empty the drain tank 

if  full

  

User manual, Sec.4 

Video 70 

71: The autoclave blocks due to use of bad water

  

Empty the water tank if necessary 

→ 

Pour in fresh demineralized or distilled water (below 3.0µS/cm) 

 

User manual, Sec.4 

Video 71 

72: The closing system was blocked as the lid was on its 

way up

 

Remove possible blocking of the lid 

User manual, Sec.4 

73: The sterilization temperature has not been 

reached fast enough during the back-flush

 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the water level is too low 

→ 

Press C to reset and then 

restart the cycle  

User manual, Sec.4 

74: The right temperature for building up pressure 

for sterilization has not been reached fast 

enough 

 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the water level is too low 

→ 

Increase the “Ambient 

sensitivity” level settings 

→ 

Press C to reset and then restart the cycle 

User manual, Sec.4 

75: The right sterilization temperature has not 

been reached fast enough

 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if the water level is too low 

→ 

Increase the 

“Ambient sensitivity” level settings 

→ 

Press C to reset and then restart the cycle 

User manual, Sec.4 

81: Low level in either the water tank or the oil 

bottle

 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank or change the oil bottle 

User manual, Sec.4 

Video 81 

84: The sterilization temperature has not been 

maintained during sterilization 

 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank if needed 

User manual, Sec.4 

Video 84 

86: The safety ring has been activated

 

Press C to reset the autoclave and then restart the machine 

→ 

Make sure that the handle is correctly placed in the holder 

User manual, Sec.4 

Video 86 

88: The steam pressure is too low during 

sterilization

 

Pour demineralized or distilled water (below 3.0µS/cm) into the water tank in case of low water level 

→ 

Press C to reset the 

autoclave and then restart the cycle 

User manual, Sec.4 

89: The steam pressure is too high during 

sterilization

 

Check that the drain tube is not blocked, broken or bent 

→ 

Empty the drain tank if full 

→ 

Press C to reset the autoclave and 

then restart the cycle  

User manual, Sec.4 

90: The pressure during pressure test was too low 

 

Check that your compressor is turned on 

→ 

Check the gasket on the lid 

→ 

Remove potential dirt 

→ 

Make sure that the lid is 

placed correctly in the lid holder  
Check if the O-ring for the filter house is mounted correctly 

→ 

Change the O-ring if damaged 

→ 

Mount the lid for the filter 

house correctly 

→ 

Replace the gasket on the lid if it is broken 

User manual, Sec.4 

Video 90 

91: The pressure fell during the pressure test

 

Check the gasket on the lid 

→ 

Remove potential dirt 

→ 

Make sure that the lid is placed correctly in the lid holder  

Check if the O-ring for the filter house is mounted correctly 

→ 

Change the O-ring if damaged 

Make sure that the lid is 

placed correctly in the holder 

→ 

Mount the lid for filter house correctly 

→ 

Replace the gasket on the lid if it is broken 

User manual, Sec.4 

Video 91 

92: The power supply to the autoclave has been 

cut out, or the C button has been pressed 
during the cycle

 

Press C to reset the machine and then restart the cycle 

User manual, Sec.4 

Video 92 

94: Oil consumption during standby

 

Disconnect power from the autoclave and wait 10 seconds 

→ 

Reconnect the autoclave to the power and then restart 

User manual, Sec.4 

95: This error can only occur if the autoclave is 

connected to an automatic water supply. The 
autoclave has “asked” for water more than 7 
times during standby

 

Disconnect power from the autoclave and wait 10 sec. 

→ 

Reconnect the autoclave to the power and then restart 

→ 

Check 

that water is being supplied to the autoclave 

→ 

Check if the autoclave has a leakage. 

User manual, Sec.4 

97: No water during external wash

 

Press C to reset the machine 

→ 

Verify that the chamber is drained. If not, empty the chamber (see section 3.5 Drain Chamber 

in the User manual) 

→ 

Restart the cycle 

User manual, Sec.4 

98: Lid did not close

 

Press C to reset and wait until the lid is in top position 

→ 

Make sure that the lid is placed correctly in the lid holder (look for 

things that might obstruct complete closure and remove these) 

→ 

Start the machine again  

User manual, Sec.4 

If the above does not solve the problem or for other possible 
displayed messages, please refer to the user manual. 

Alternatively, please contact your authorized supplier. 

Check out tech forum videos  
on our YouTube channel 

REF 6478270 

510263-03.1B_EN 

Learn More 

DAC UNIVERSAL 

 

USER GUIDE FOR DISPLAYED MESSAGES

 

Summary of Contents for DAC UNIVERSAL

Page 1: ...al Sec 4 Video 86 88 The steam pressure is too low during sterilization Pour demineralized or distilled water below 3 0 S cm into the water tank in case of low water level Press C to reset the autocla...

Page 2: ...n asennettu oikein pidikkeeseen Kayttoohje Osio 4 Video 86 88 H yrynpaine on liian alhainen steriloinnin aikana Kaada vesisailioon demineralisoitua tai tislattua vetta alle 3 0 S cm jos vesitaso on al...

Page 3: ...correct in de houder is geplaatst Gebruikershandleiding Deel 4 Video 86 88 De stoomdruk is te laag tijdens sterilisatie Giet gedemineraliseerd of gedestilleerd water lager dan 3 0mS cm in de watertank...

Page 4: ...ts Tryck pa C for att aterstalla autoklaven och starta sedan om maskinen Se till att handtaget ar korrekt placerat i hallaren Anv nderhandbok Avs 4 Video 86 88 ngtrycket r f r l gt under sterilisering...

Page 5: ...4 73 C 4 74 3 0m Ambient sensitivity C 4 75 3 0m Ambient sensitivity C 4 81 3 0m 4 81 84 3 0m 4 84 86 C 4 86 88 3 0m C 4 89 C 4 90 DAC UNIVERSAL 5 8 4 90 91 4 91 92 C C 4 92 94 10 4 95 7 4 97 C 4 98...

Page 6: ...sede Istuzioni d uso Sez 4 Video 86 88 Pressione del vapore troppo bassa durante la sterilizzazione Versare acqua demineralizzata o distillata con massimo 3 0 S cm nel serbatoio se il livello risulta...

Page 7: ...a correcta en el soporte Manual del operador Sec 4 Video 86 88 La presi n es demasiado baja durante la esterilizaci n En caso de que el nivel del agua sea bajo a ada agua destilada o desmineralizada m...

Page 8: ...t Tryk C for at nulstille maskinen og genstart derefter Brugermanual Afs 4 89 Damptrykket er for h jt under sterilisationen Tjek at dr nslangen ikke er blokeret delagt eller b jet T m dr ntanken hvis...

Page 9: ...o 86 88 La pression de la vapeur est trop basse lors de la st rilisation Versez de l eau d min ralis e ou distill e sous 3 0mS cm dans le r servoir d eau en cas de faible niveau d eau Appuyez sur C po...

Page 10: ...r Sterilisation nicht eingehalten F llen Sie den Wasserbeh lter nach Bedarf mit entmineralisiertem oder destilliertem Wasser auf unter 3 0mS cm Gebrauchsanweisung Abschnitt 4 Video 84 86 Der Sicherhei...

Page 11: ...64 C 4 67 3 4 70 4 70 71 3 0 S cm 4 71 72 4 73 3 0 S cm C 4 74 3 0 S cm Ambient sensitivity C 4 75 3 0 S cm Ambient sensitivity C 4 81 3 0 S cm 4 81 84 3 0 S cm 4 84 86 C 4 86 88 3 0 S cm C 4 89 C 4...

Reviews: