61 25 665 D3437
100
D3437
.
076.01.15
.
02 06.2012
5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
E1 10 23/24/25/26 GALILEOS
Error code
Description
Actions required
see
E1 10 23
E1 10 24
E1 10 25
E1 10 26
Board DX89 does not have valid
data via the X-ray detector.
The X-ray detector has been
replaced. Board DX89 does not
have valid data via the X-ray
detector.
This error message should not
occur in the application.
Board DX89 was replaced and must
be registered in the system.
The X-ray detector has not been
initialized. Board DX89 does not
have valid data via the X-ray
detector.
This error message should not
occur in the application.
● Perform service routine S009.7 (copy data
from DX11 to board DX89).
● Replace the X-ray detector.
Please report this event to the Customer
Service Center to help us improve the
product.
● Perform service routine S009.7 (copy data
from DX11 to board DX89).
● Replace the X-ray detector.
Please report this event to the Customer
Service Center to help us improve the
product.