61 25 665 D3437
108
D3437
.
076.01.15
.
02 06.2012
5 Messages
Sirona Dental Systems GmbH
5.5 List of error messages
Service Manual GALILEOS
E6 41 06
E6 41 07
E6 41 08
E6 41 09
Error code
Description
Actions required
see
E6 41 06
Module failed in TTP (time trigger
protocol) (detected on master side).
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
● Check the CAN bus.
● Replace board DX41.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
S. [ → 128],
S. [ → 343],
S. [ → 61]
Error code
Description
Actions required
see
E6 41 07
TTP (time trigger protocol) timeout
error (detected on slave side)
The module was temporarily not
addressed by the master:
Undervoltage on the master side
Procedure error in the software
Master (DX11) receives no return
commanding from the module
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
● Check the CAN bus.
● Check power supply of board DX11;
measuring point 3.3 V on board DX1 (see
wiring diagrams).
If 3.3 V is present ...
● Replace board DX11.
If 3.3 V is not present ...
● Replace board DX1.
Error code
Description
Actions required
see
E6 41 08
General fault detected locally on
module (slave side). CAN controller
being reinitialized.
● Check the CAN bus.
● Check software versions on the info
screen or by running service routine
S008.2, perform software update if
necessary.
● Replace board DX41.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
S. [ → 128],
S. [ → 238],
S. [ → 61],
S. [ → 61]
Error code
Description
Actions required
see
E6 41 09
Board DX41 is installed in the unit,
but not configured.
● Perform device configuration via service
routine S017.9.
Summary of Contents for galileos
Page 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Page 2: ......
Page 387: ......