Service Manual GALILEOS
5.5
List of error messages
båÖäáëÜ
E6 89 08
E7 89 10
E7 89 12
E5 89 13
E6 89 20
Error code
Description
Actions required
see
E6 89 08
General fault detected locally on
module (slave side). CAN controller
being reinitialized.
● Check the CAN bus.
● Check software versions on the info
screen or by running service routine
S008.2, perform software update if
necessary.
● Check board DX89, replace if necessary.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
S. [ → 128],
S. [ → 238],
S. [ → 61],
S. [ → 134],
S. [ → 343]
Error code
Description
Actions required
see
E7 89 10
Module is stuck in bootloader stage. ● Check operating status of board (note
LED states).
If the board remains in the bootloader stage ...
● Run software update.
● Check board DX89, replace if necessary.
S. [ → 61],
S. [ → 134],
S. [ → 343]
Error code
Description
Actions required
see
E7 89 12
Unit is not ready for operation
This error is a sequential fault.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and check function.
If the error occurs repeatedly...
● Repeat procedure and observe causal
error messages.
Error code
Description
Actions required
see
E5 89 13
Error when writing to EEPROM
IMPORTANT: Stored data may be
lost.
● Acknowledge error and repeat procedure.
● Run software update.
If the error occurs again ...
● Check log memory (via extended details).
● Check x-ray detector, replace if
necessary.
S.,
S. [ → 319]
Error code
Description
Actions required
see
E6 89 20
Faulty voltage supply of DX89.
● Check cable L13, replace if necessary.