61 25 665 D3437
126
D3437
.
076.01.15
.
02 06.2012
6 Troubleshooting
Sirona Dental Systems GmbH
6.1 Error logging memory
Service Manual GALILEOS
6.1.1
Example of error logging data
Example of error logging data
System time
2006-03-06, 20:13:02
System time (clock on DX11)
Entry type
[Message]
General system event
[M val]
General system event with additional
value
[Error]
Error event
[DeviceError]
Data for error event on a module
[Error Sidexis]
Network error event
[Stringname]
Free status texts
[Stringsegment]
Additional data (string name)
[RTC Date / Time Change]
Date and time of a SIDEXIS PC
[PC Date / Time]
Date and time of the DX11 set
[Compression table]
Compression table
Entry data [Message] Self-test: Successful
Self-test completed successfully
Recording started
Start of a recording
Value: 9000
Sequence ID of the recording
Recording stopped
End of an exposure
Recording cancelled
Exposure cancelation
Termination state
Reason for ending exposure
Value: 0
Exposure completed
Value: 1
Exposure cancelation by user
Value: 2
Exposure cancelation due to internal
error
Imagetransfer started
Start of image data transfer
Imagetransfer stopped
End of image data transfer
Logbook started
Corresponds to unit switch-on
Image state switched to Released
Exposure has been delivered to
SIDEXIS XG and confirmed by SIDEXIS
XG.
Other entry data which document the occurrence of a rescue event include:
● Image state switched to Rescue
● Rescue request Sidexis Error
● Rescue request Sidexis TrackEpilogue
● Rescue request Sidexis Timeout
These entry data may also occur after "Recording stopped" or "Cancel" and
indicate an exceptional circumstance. You can supply important information for
error diagnosis in coordination with the Sirona Customer Service Center.
Entry data [Error]
E6 07 06
Error code
ERR_DX7_TTP_LOST
Clear text error display
Summary of Contents for galileos