61 25 665 D3437
152
D3437
.
076.01.15
.
02 06.2012
6 Troubleshooting
Sirona Dental Systems GmbH
6.9 Fault diagnosis of the X-ray detector and on board DX89
Service Manual GALILEOS
6.9.1
LEDs on board DX89
LEDs on board DX89
The diodes PLL_FPGA, PLL_CCD as well as (A) (image memory test)
and (D) (gettering) in particular should be observed (LED statuses and
their significance in case of an error [ → 153]).
A
Image memory test
B
Operating voltage on board DX89
C
Operating voltage on image amplifier
D
Gettering