Adjusting/calibrating the unit
Service Manual GALILEOS
7.1
General information about unit adjustment and calibration
båÖäáëÜ
7.1
General information about unit adjustment and
calibration
General information about unit adjustment and calibration
Checking the mechanical unit adjustment: GALILEOS
Start by checking the mechanical unit adjustment. This step is a
prerequisite for the subsequent adjustment and calibration of the unit.
Please adhere to the following order when adjusting and calibrating the
system:
GALILEOS calibration procedure
● Diaphragm exposure: "Type 1/Type 2" [ → 167] diaphragm or
diaphragm exposure: "Type 3" diaphragm [ → 171]
● Checking the radiation field [ → 178]
● Dosimetry [ → 179]
● Sensor adjustment [ → 181]
● Iris adjustment [ → 182]
● Shading calibration [ → 183]
● Distortion calibration [ → 185]
● Geometry calibration [ → 187])
Tip: It may be helpful to use the SIDEXIS XG coloring function to evaluate
the image.