7 Adjusting/calibrating the unit
Sirona Dental Systems GmbH
7.1 General information about unit adjustment and calibration
Service Manual GALILEOS
7.1.6.2
Geometry phantom
Geometry phantom
IMPORTANT: Make sure that the phantom is securely fastened and in an
upright position in the bite block holder of the unit.
For the geometry calibration [ → 187], you must insert the geometry
phantom in the block holder of the unit.
This phantom must be removed again for all other calibration steps.
Up to unit serial number 8,699
(GALILEOS Comfort) and 48,699
(GALILEOS Compact)
IMPORTANT
This geometry phantom is not suitable for calibration with Facescan.