61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
165
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.1
General information about unit adjustment and calibration
båÖäáëÜ
From unit serial number 8,700
(GALILEOS Comfort) and 48,700
(GALILEOS Compact)
➢ Insert the geometry phantom (A) into the pan bite block holder (B) on
the unit and secure it with the screw (C).
IMPORTANT
Once it has been inserted into the bite block holder, the geometry
phantom must be aligned vertically and horizontally with the spirit level,
so that calibration can be performed correctly.