61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
167
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
7.3
Adjustment and calibration via the "Service
Functions" menu
Adjustment and calibration via the "Service Functions" menu
7.3.1
Diaphragm image
Diaphragm image
7.3.1.1
"Type 1/Type 2" diaphragm
"Type 1/Type 2" diaphragm
✔
The
"Service functions"
menu is selected.
1. Click on the
"Image acquisition"
button.
ª
SIDEXIS XG makes the unit ready for exposure [ → 162].
ª
Service routine S030.5 is displayed on the control panel.
2. Take an exposure (85 kV/21 mAs) [ → 162].
3.
For "GALILEOS Comfort":
Evaluate the image.
Summary of Contents for galileos