61 25 665 D3437
172
D3437
.
076.01.15
.
02 06.2012
7 Adjusting/calibrating the unit
Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu
Service Manual GALILEOS
ª
The brightness distribution along the border surrounding the
image on all sides must be uniform (A).
ª
The distance between the bottom edge and the lowest point in
the image should be 30 ± 5 pixels (measure with SIDEXIS scale)
(B).
If the distance between the bottom edge and the lowest point in
the image is out of tolerance (E) or the brightness distribution
along the surrounding border is not uniform (C), the diaphragm
must be adjusted mechanically [ → 197].
5.
For "GALILEOS Compact":
Evaluate the image.
A+B
Adjustment OK
C
Adjustment not OK
E
Permissible tolerance: 30 pixels ± 5 pixels
Summary of Contents for galileos
Page 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Page 2: ......
Page 387: ......