61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
175
Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
ª
The upper edge of the lower lead diaphragm must be within
tolerance, i.e. lie inside of the auxiliary lies (A).
If the edge is out of tolerance (B), the diaphragm must be
adjusted mechanically [ → 197].
5. If the exposure is OK (A), confirm this by clicking the check box
located to the right of the
"Upper jaw"
button.
ª
The box will appear checked.
ª
The adjustment for the "Upper jaw" diaphragm setting is now
complete.
ª
The
"Lower jaw"
button is selectable.
6. Continue the calibration procedure with the adjustment of the "Lower
jaw" diaphragm setting.
A
Adjustment OK
B
Adjustment not OK