7 Adjusting/calibrating the unit
Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu
Service Manual GALILEOS
Adjustment of the "Lower jaw"
diaphragm setting
✔
The
"Service functions"
✔
The
"Diaphragm"
tab is selected.
✔
The
"Lower jaw"
is selectable in the
"Image acquisition"
menu area.
1. Set the rotary knob on the tube assembly to the "lower jaw" position.
2. Click on the
"Lower jaw"
button.
ª
SIDEXIS XG makes the unit ready for exposure [ → 162].
ª
Service routine S030.5 is displayed on the control panel.
3. Take an exposure (85 kV/0.21 mAs) [ → 162].
4. Evaluate the image.