7 Adjusting/calibrating the unit
Sirona Dental Systems GmbH
7.3 Adjustment and calibration via the "Service Functions" menu
Service Manual GALILEOS
ª
The shading exposure and the evaluation of the shading
calibration is displayed.
If the information in the message window indicates that
calibration is
not OK and/or not possible
, keep repeating the
procedure starting with Step b) until calibration is OK and/or
possible.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
IMPORTANT: No foreign bodies may be visible on the shading
exposure. If this is the case, check the beam path for foreign
bodies, remove them if necessary and repeat the calibration.
5. If the calibration is OK or possible, click the
"Save values"
button.
ª
The calibration is saved.
ª
The shading calibration is now complete.