61 25 665 D3437
194
D3437
.
076.01.15
.
02 06.2012
7 Adjusting/calibrating the unit
Sirona Dental Systems GmbH
7.6 Mechanical adjustments
Service Manual GALILEOS
4.
Move the ring center to the left or right:
NOTICE! Do not undo the screws completely!
Loosen the four screws (C) slightly.
5. Correct the position carefully and then retighten the screws.
If the center of the ring cannot be fully adjusted using the screws (C)
then proceed with adjusting the swivel arm. Otherwise, the
mechanical adjustment is now complete and you may begin
calibration.
6. Re-attach the "arm cover".