61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
219
Sirona Dental Systems GmbH
8
Service
Service Manual GALILEOS
8.3
Basic operating procedures in the service menu
båÖäáëÜ
8.3.4
Saving parameters
Saving parameters
Once one or a number of parameters have been selected via a service
routine, the current selection must be saved so that it is applied in the unit.
Title: Saving parameters in GALILEOS
Easypad touchscreen
Example
You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔
The Memory key (C) lights up.
1. Touch the Memory key (C).
ª
The R key (D) lights up.
2. Touch the R key (D).
ª
The selected setting is saved to non-volatile memory.
Saving parameters in GALILEOS
Multipad
Example
You want to run service routine S017.6 to save the selected option
"Remote control enabled".
✔
The LED above Memory key (C) lights up.
1. Press Memory key (C).
ª
The LED above the R key lights up.
2. Press R key (D).
ª
The selected setting is saved to non-volatile memory.