227
Sirona Dental Systems GmbH
8
Service
Service Manual GALILEOS
8.5
S005: General X-ray tube assembly service
båÖäáëÜ
8.5.4
S005: Test step 8
S005: Test step 8
Automatic adjustment of pulse preheating
1. Call service routine S005.8 [ → 213].
ª
An inactive progress indicator in selection field 1 and the
message "FFFF" in selection field 2 signal that the system is
ready for compensation.
2. Start the automatic adjustment by pressing and holding the release
button.
IMPORTANT: Keep pressing the release button until adjustment is
completed and the new offset value for preheating is displayed. If you
interrupt the adjustment procedure prematurely by letting go of the
release button, the message "EEEE" appears in selection field 2. This
message must be acknowledged by pressing the R key.
ª
When pressing the release button, radiation is released for 2 s to
warm up the tube assembly to operating temperature. This is
followed by the automatic tuning routine.