8 Service
Sirona Dental Systems GmbH
8.6 S007: Error logging memory
Service Manual GALILEOS
3. Use the arrow keys (A) in selection field 1 to select the required error
event (66 in the example).
ª
The corresponding error message is displayed in selection field
2 [ → 82].
ª
Selection field 3 displays the date and time of the error event.
4. Exit the service routine [ → 220].
S007_1 GALILEOS
Multipad
1. Call service routine S007.1 [ → 213].
2. Use the arrow keys (A) in selection field 1 to select the required error
event (66 in the example).
Use the patient symbol keys to set the increment for scrolling
between the error numbers:
Patient symbol key 1 (left) = increment 1 (factory setting)
Patient symbol key 2 = increment 10
Patient symbol key 3 = increment 100
The LED above the selected patient symbol key is lit up.