61 25 665 D3437
240
D3437
.
076.01.15
.
02 06.2012
8 Service
Sirona Dental Systems GmbH
8.7 S008: Update service
Service Manual GALILEOS
8.7.2
S008: Test step 3
S008: Test step 3
Confirming the unit serial number
1. Call service routine S008.3 [ → 213].
2. Confirm the serial number displayed by pressing the R key (A).
3. Exit the service routine [ → 220].
IMPORTANT
If the backup copy of the old unit serial number does not match the new
one after replacing a module, the entry of the serial number is activated.
If an incorrect serial number is entered, the message "FFFF" appears
on the display. In this case, the service routine can be run again.
Summary of Contents for galileos
Page 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Page 2: ......
Page 387: ......