61 25 665 D3437
248
D3437
.
076.01.15
.
02 06.2012
8 Service
Sirona Dental Systems GmbH
8.9 S011: Dosimetry (without ring movement)
Service Manual GALILEOS
2. Initiate the radiation.
ª
Radiation uses 200 pulses and 85kV/42mAs.
IMPORTANT: If you let go of the release button before the
maximum radiation time has elapsed, radiation is terminated
prematurely and the exposure is interrupted. The actual radiation
time is
not
displayed.
If you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the
header of the Easypad or in selection field 2 on the Multipad
(automatic exposure blocking).
3. Exit the service routine [ → 220].
8.9.2
S011: Test step 9
S011: Test step 9
Current measurement (unpulsed)
1. Call service routine S011.9 [ → 213].
ª
Selection field 1 displays the kVmA level, and selection field 2
displays the maximum radiation time.
The kVmA level and the maximum radiation time are preset and
cannot be changed.
2. Initiate the radiation.
ª
The maximum set radiation time has elapsed.
IMPORTANT: If you let go of the release button before the
maximum radiation time has elapsed, radiation is terminated
prematurely and the exposure is interrupted. The actual radiation
time is
not
displayed.
If you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the
header of the Easypad or in selection field 2 on the Multipad
(automatic exposure blocking).