61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
263
Sirona Dental Systems GmbH
8
Service
Service Manual GALILEOS
8.11
S017: Configuration service
båÖäáëÜ
8.11.7
S017: Test step 9
S017: Test step 9
Activate/deactivate operation with board DX41
This service routine is used to configure operation of the system with or
without module DX41. This configuration is necessary for software
updates and module replacement of board DX11 if systems with or
without board DX41 should be supported.
* Factory setting
1. Call service routine S017.9 [ → 213].
ª
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
Selection field
Code
Function
1
00
Board DX41 inactive*
01
Board DX41 active