Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.2
Height adjustment motor (M1_4)/spindle
båÖäáëÜ
6. Unscrew the remaining screws from the protective plates (E).
7. Remove the covering plates (E) (top and bottom) from the connection
box (F) of board DX32.
8. Cable L3 can stay on the top covering plate (E) (remove any possible
shield terminal).
Tip: The covering plate can be folded away simply at the sides and
put away in the stand sideways (to protect against scratches suitable
padding should be put in between.).
9. Remove connector X2 from board DX32 and remove the protective
conductor.
10. Remove cable L2 from terminal X100 and pull it downwards from out
of the connection box (F).
11. Remove connector X1 from board DX32.
12. Loosen the two left-hand screws (G).
13. Loosen the two screws on the right (I) and remove the connection box
including board DX32.
L3
E
L3
G
DX32
X100
L2
X1
F
H
X2
I