Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.2
Height adjustment motor (M1_4)/spindle
båÖäáëÜ
Installing the new motor
Install the height adjustment motor in the reverse order of removal.
Please observe the following:
Nuts: When fastening the motor, make sure that all three screws are
tightened uniformly and protrude approx. 3 mm out of the nut.
Acorn nuts: If acorn nuts have been installed in the unit, turn the acorn
nuts to the end stop.
CAUTION! Do not forget to reattach all connectors or cables, route them
in their original position and reattach all cable ties and cable clamps.
Make sure that none of the cables are crushed by the cover plates of the
DX32 connection box.
Final work
With the "height adjustment" service kit
1. After reinstalling the spindle, screw the first of the two screws (E)
back into the spindle holder.
2. Then remove the locking pin and screw in the second of the two
screws (E).
3. Attach the cover (D).
4. Remove the clamp (A).
5. Only then should you check the travel function of the slide.
Without the "height adjustment" service kit
1. Attach the spindle holder (F) with the two screws (E).
2. Attach the cover (D).
3. Reattach the upper limit stop to the previously marked position.
4. Only then should you check the travel function of the slide.