61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
315
Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.6
X-ray tube unit
båÖäáëÜ
Installing the diaphragm unit
➢ NOTICE! Do not forget to re-install the filters or in the case of "Type
1" diaphragm unit, the primary diaphragm.
The diaphragm is installed in the reverse order to removal.
With the "Type 3" diaphragm:
Attaching the covers
➢ Reattach the covers.
With the "Type 3" diaphragm:
➢ NOTICE! Do not force on the adjusting knob. Make sure the locking
function of the adjusting knob works properly when setting the button
in place.
Set the adjusting knob in place with the silicone ring.
Updating the ID label
➢ Update the nameplate at the tube assembly cover.
To do so, affix the supplied label as shown in the figure.
IMPORTANT
Note the order for installing the filters!
The image quality will be impaired if the filters are installed in the wrong
sequence. Insert the aluminum filter (D) first, followed by the copper filter
(M).
Summary of Contents for galileos
Page 1: ...MSKOMNO d ifiblp kÉï ë çÑW båÖäáëÜ pÉêîáÅÉ j åì ä GALILEOS Service Manual ...
Page 2: ......
Page 387: ......