Service Manual GALILEOS
9.6
X-ray tube unit
båÖäáëÜ
9.6.3
What has to be done after replacing the X-ray tube
assembly?
What has to be done after replacing the X-ray tube assembly?
1. Perform a complete unit adjustment or calibration [ → 155].
2. Perform an acceptance test (for Germany only) without calling in an
expert.
A
Cable L5 → Socket J6 on board DX6
B
Cable L6 → Socket J2/J3 on board DX6
C
Grounding cable → Connector X304 on board DX6
D
Cable L3 → Connector X3 on board DX6
E
Laying cables correctly on the cover plate
F
Cable L15 → Socket J5 on board DX6
G
Cable routed on left side of tube assembly:
2x L21 and L20
H
Cable L12 routed on right side of tube assembly.
IMPORTANT
Since board DX6 is also replaced with the tube assembly, you MUST
also follow the instructions in the chapter entitled Measures following