61 25 665 D3437
D3437
.
076.01.15
.
02 06.2012
319
Sirona Dental Systems GmbH
9
Repair
Service Manual GALILEOS
9.8
X-ray detector
båÖäáëÜ
9.8
X-ray detector
X-ray detector
9.8.1
Replace X-ray detector
Replace X-ray detector
Removing the covers
➢ Remove the "x-ray detector cover".
Remove cable
1. CAUTION! Risk of injury! The cover plate may have sharp edges.
Carefully pull cover plate (A) upwards to remove it from the X-ray
detector.
2. Loosen the screws (B) and (C), as well as clamps (D) and (E).
3. Remove cable L13 from connector X201 (F) on board DX89.
IMPORTANT
For Facescan units:
The FaceScan must be removed [ → 325] from units with
FacescanFacescan fitted before the X-ray detector can be replaced.