61 25 665 D3437
320
D3437
.
076.01.15
.
02 06.2012
9 Repair
Sirona Dental Systems GmbH
9.8 X-ray detector
Service Manual GALILEOS
4. NOTICE! Make sure that the grounding cable does not slip into the
ring. Secure it with a cable tie or piece of adhesive tape if necessary.
Loosen screw (G) and disconnect the grounding cable (H).
Removing the X-ray detector
➢ CAUTION! The x-ray detector is heavy!
Loosen the screw (J), swing the X-ray detector slightly upwards and
lift it out of the holder on the ring.
IMPORTANT: Depending on the unit hardware version involved,
there may be a second screw located on the side opposite screw (J).
If so, this screw(*) must be loosened in order to remove the X-ray
detector. This second screw does not have to be used during
reassembly.