61 25 665 D3437
328
D3437
.
076.01.15
.
02 06.2012
9 Repair
Sirona Dental Systems GmbH
9.9 Facescan
Service Manual GALILEOS
1. Hang the scan unit from behind in the inlets of the X-ray detector.
2. Attach the scan unit loosely with 6 screws (D) - do not screw tightly!
3. Mount the installation aid (C) from below on both sides onto the first
cooling vent (H) of the face scan unit.
4. Insert the fastening element (F) into the installation aid (C).
5. Turn the holding plates (G) over the chassis plate of the face scan
unit.
6. Push the installation aid (C) with the scan unit on to the sensor
surface (H) of the X-ray detector.
7. Align the installation aid (C) to the center of the sensor surface (H).
IMPORTANT
Possible faulty alignment
Displacement of the installation aid (C) can be caused through further
fastening of the scan unit with the safety screws (E).
➢ When screwing in the safety screws (E), please ensure that the
installation aid (C) always remains aligned.