9 Repair
Sirona Dental Systems GmbH
9.10 Head fixation device
Service Manual GALILEOS
9.10
Head fixation device
Head fixation device
9.10.1
Replacing receptacle element for head fixation (for unit
with head fixation device)
Replacing receptacle element for head fixation (for unit with head fixation device)
Remove defective receptacle element.
1. Using the Up/Down buttons on the control panel, switch the device on
and move it to a comfortable working height to remove the acquisition
unit.
2. Remove the head fixation device (see operating instructions).
3.
If the bore hole (C) on the defective acquisition unit is not available:
Press the locking button (A) and move the flange (B) forward to
expose the bore hole (C).
4. Loosen screw (D) and remove the defective acquisition unit.
Install the new acquisition unit.
1.
If the bore hole C on the new acquisition unit is not available:
Press the locking button (A) and move the flange (B) forward to
expose the bore hole (C).