Repair
Service Manual GALILEOS
9.10
Head fixation device
båÖäáëÜ
2. IMPORTANT: Do not tighten the screw yet. It should not be possible
to rotate the acquisition unit.
Screw the new receptacle element on to the unit with screw (D) so
that the laser localizer (E) is facing forward.
3. Push the head fixation device (F) into the acquisition unit (see
operating instructions).
4. Press the locking button (A) and push the flange (B) including the
head fixation device backwards so that the light localizer (E) is
exposed.
Adjust the light localizer.
Adjusting the light localizer (GALILEOS)
1. CAUTION! Keep a minimum distance of 100 mm between the eye
and the laser. Do not look directly into the laser beam.
Switch the laser light on using the light localizer button on the control
panel.